Preparation of Reagents:
Norml Goat Serum Blocking Solution
20 ml normal goat serum (Filter sterilize with 0.2 µM filter)
Volume to 100 ml with PBS or TBS.
Store 2-8 °C, discard after 3 months.
IHC Antibody Diluent (50mM TBS, 1% BSA)
190 ml 1X TBS
2 g BSA (Sigma A9647)
Volume to 200 ml with TBS.
Store 2-8 °C, discard after 1 year.
*May want to use a perservative.
Wash solution
Can be any of the following: PBS, TBS or 0.01% Tween 20 in dH20. TBS wash solution is recommended for phospho antibodies.
10% Triton-X 100 Stock
10 ml Triton-X 100
90 ml dH20
0.25% Triton-X 100
25 µl 10% Triton-X 100
1 ml TBS
Secondary Antibody
Goat anti-Rabbit IgG Antibody-Dylight conjugated. (Bethyl cat# A120-101D2, A120-101D3, A120-101D4, or cross-adsorbed antibodies A120-201D2, A120-201D3, A120-201D4) Dilute 1:100 with IHC antibody diluent. Prepare just prior to use. Protect from light. Optimal working dilutions should be determined experimentally by the investigator.
Procedure:
- For cells in chambered microscope slides or cells grown on coverslips: Gently rinse off media with TBS - 3 changes for 1 minute each. For cytospins: Allow to air dry for 1-2 minutes after preparation.
- Formaldehyde fixation - 10-30 minutes
- Wash Solution - 3 changes for 5 minutes each
- Circle cytospin with a hydrophobic barrier pen. Do not allow cells to dry for the remaining procedure.
- Permeabilization step: 0.25% Triton-X 100 - 2-10 minutes. Optimal working dilution and incubation time should be determined experimentally by the investigator.
- Wash Solution - 3 changes for 5 minutes each
- Normal goat serum Blocking Solution - 15 minutes
- Primary Antibody Incubation: 30 minutes - room temperature. Prepare primary antibody with IHC Antibody Diluent. Optimal working dilutions should be determined experimentally by the investigator.
- Wash Solution - 3 changes for 5 minutes each
- Secondary Antibody Incubation: 30 minutes - room temperature. Protect from light.
- Wash Solution - 3 changes for 5 minutes each
- Remove chamber from microscope slide.
- Mount with fluorescent mounting media and coverslip. Use fluorescent mounting media with DAPI if counterstaining is desired.
NOTE: This procedure is suitable for Phospho-specific Antibodies